Quick Specifications

Sensor Inputs
RF Channels 1 or 2
RF Frequency Range 9.9 kHz to 110 GHz*
Peak Pwr range -50 to +20 dBm*
CW Pwr range -70 to +44 dBm*
   
Acquisition and Measurement System
Time resolution 0.2 ns
A/D Converter 14 bit
DSP 32 bit floating point
   
Trigger
Ext Trig range, impedance ±5V, 1 Mohm
Min trig pulse width 15 ns
Max trig rate 30 MHz
   
Calibrator Source
Internal Calibrator 50.025 MHz, CW -60 to +20 dBm
* Sensor Dependent, Calibrator Dependent
 
Pulse Mode Operation (Automatic)
Pulse width Pulse rise-time
Pulse fall-time Pulse period
Pulse repetition frequency Pulse duty cycle
Pulse off-time Peak power
Pulse “on” power Pulse overshoot (dB or %)
Waveform Average power Top level power (IEEE spec)
Bottom level power (IEEE spec) Edge delay
Edge skew (2 channel instruments only)
 
Marker Measurements
Markers (vertical cursors)
Settable in time relative to the trigger position
Markers independently
Average, minimum, peak power at a single time offset
Pair of Markers
Average, minimum, peak power over the interval between markers, power ratio between markers

Options

4541  Peak Power Analyzer, single channel, front panel inputs.
4542 Dual channel, front panel inputs
-02 Rear sensor inputs
-03 Calibrator, rear panel output
-30 Warranty extended to 3 years